TURBICUBE

|
|
TURBICUBE, In line turbidity transmitter
- Flow-through in-line turbidimeter acc. ISO 7027 - Turbidity transmitter - With two IR emitters and one IR receiver - Two models: 0.01...20 FNU (5 ranges) 0.1...1000 FNU (5 ranges) - 4...20mA output signal (4 wires) - Flange or thread connection |
| The TURBICUBE measures the turbidity of liquids using the light absorptionand scattered light measuring method. The inline measuring cell includes two IR emitters and one IR receiver inpulsating light mode (IR measuring method according ISO 7027). When interconnected, the BAMOPHOX TUR, measuring amplifier, analysesthe 4 to 20 mA signal generated by the TURBICUBE. |
The flow-through cell allows the turbidity measurement directly on the line,without sampling system. Its built-in transmitter is convenient for transmittingthe signal to installed DCS or PLC or BAMOPHOX TUR. The 4-20 mAcorresponds to the DIP Switch selectable range:
TURBICUBE 20: from 0.01 to 20 FNU in 5 selectable ranges TURBICUBE 1000: from 0.1 to 1000 FNU in 5 selectable ranges |
• Drinkable and waste water plants, well water monitoring • Filtration control, flocculants dosage, phase separation • Pollution survey |
Distilled water:0 FNU Filtered apple juice:2...4 FNU River water:1...4 FNU Milk (fat 3%):around 4000 FNU Filtered Pale ale: 1...6 FNU Raw beer: 20...400 FNU Distilled water, 10% milk:around 20 FNU |
Terminal housing: PBT, IP65 Auxiliary power supply: 24V DC (10...30 V DC); connected load 0.5 W. Note: connecting also the measuring circuit to the auxiliary power supply of the load, it increases of 24 V DC x 20mA = 0,48W Power consumption: 0,6 W Output signal: 4-20 mA for connection to the BAMOPHOX TUR 436. Note: if a different measuring amplifier is used, an external 10 to 30 V DC voltage source is required. Ambient temperature: + 5...+ 45°C Process temperature: + 5...+ 60°C Accuracy: ± 5% reading, ± 1% full scale in use Maximum pressure: 6 bar at 20°C / 1 bar at 60°C Status indication: correct is a green LED; failure is a red LED Selectable ranges TURBICUBE 1000 :from 0,1 to 50 / ...100 / ...200 / ...500 (300*) FNU from 100 to 1000 FNU * fixtures with DN80/DN100 Selectable ranges TURBICUBE 20:from 0,01 to 1 / ...2 / ...5 / ...10 / ...20 FNU Wetted parts:Plane wall measuring cell from polymers (PCV or PP) and anti-fouling ETR-glasses. Sealing: EPDM (FPM as an option) Process connections: from DN 20 to DN 100, flanges or union socket ends
CE Mark:In accordance with low-voltage directive (73/23/EWG), EMC directives (89/336/EEC) |
|
 |